Review: Teledyne LeCroy WaveAce 1002 – Digital Storage Oscilloscope

Low-cost oscilloscope are a booming market. With the low-cost, entry level WaveAce series Teledyne LeCroy wants to get their piece of the big cake. This review is going to have a closer look at the Teledyne LeCroy WaveAce 1002.

The WaveAce series oscilloscopes from Teledyne LeCroy are LeCroy’s answer to low-cost entry level oscilloscopes. They are available as 2- or 4-channel versions and with different bandwidths (40 – 300 MHz). LeCroy’s representative for my region, T&M Instruments [1], shipped a LecCroy WaveAce 1002 (2-channel, 60 MHz) my way to have a look at it. Let’s check it out!

The WaveAce 1002 is a 2-channel DSO with 60 MHz bandwidth, a sample rate of 1 GS/s with just one channel in use or 500 MS/s if both channels are in use, and a memory depth of 2 Mpt/Ch with just one channel in use or 1 Mpt/Ch if both channels are used. The display of the scope is a 7.0 TFT-LCD with a screen resolution of 480 x 234 pixels (QVGA).

For more details and tech-specs, visit the manufacturer’s webpage for the AceWave series [2] and the WaveAce 1002 [3].

The scope makes a very solid impression. The menus are self-explanatory and very well structured. I didn’t need to look at the manual or the built-in help functions at any point to find exactly what I was looking for.

Close-up view of the front panel. The front panel looks very clean and neatly organized.

Close-up view of the front panel. The front panel looks very clean and neatly organized.

I particularly like that LeCroy didn’t simply reinvent the wheel and actually added some value to their entry-level scopes. The WaveAce scopes come with neat functions that are usually in the domain of upper intermediate level scopes.

One of those great features is a mask testing feature. A mask of a signal with acceptable tolerance can be loaded into the scope. The scope then compares the input signal with the mask and indicates whether or not the signal is within the mask or not.

The integrated pass / fail test function makes the WaveAce series oscilloscopes perfect for quality control applications.

The integrated pass / fail test function makes the WaveAce series oscilloscopes perfect for quality control applications.

Not only does the scope give a visual pass / fail indication, there’s even a dedicated pass / fail BNC connector on the back of the scope. That way, the WaveAce scope can easily be used in automated testing quality control set-ups. The BNC’s output voltage can be selected to be active on fail or active on pass.

Connectors on the back: Pass/Fail, RS232, USB

Connectors on the back: Pass/Fail, RS232, USB

There’s no special software needed to set up the mask. The oscilloscope has an extra menu dedicated to creating a mask.

Mask can conveniently be set up using the mask generation tools of the scope itself. No external software is needed.

Mask can conveniently be set up using the mask generation tools of the scope itself. No external software is needed.

Another great feature that comes in handy is the zoom function. A common problem with a regular oscilloscope is that one can either set the time base to get a good overview of the whole signal or one can zoom into the signal but usually not both at the same time. LeCroy thought about this problem and implemented a zoom function that enables one to zoom into a section of the signal using a split screen feature without losing the original view of the signal. Sounds confusing? Take a look:

Random noise from a noise source. Top: Full captured signal, Bottom: Zoomed in on selected region of the signal

Random noise from a noise source. Top: Full captured signal, Bottom: Zoomed in on selected region of the signal

A similar split screen can be used while viewing a Fast Fourier Transform (FFT) of a signal. A mixed domain view allows the signal to be viewed in the time domain (top) and in the frequency domain (bottom) at the same time.

A bunch of noise from a noise generator. Top: View in the time domain, Bottom: Simultaneous view in the frequency domain

A bunch of noise from a noise generator. Top: View in the time domain, Bottom: Simultaneous view in the frequency domain

All screenshots I am using in this review have been generated by the scope itself, by the way. All one needs to do to save a screenshot on an external storage device is to plug the storage device into the USB port on the front panel of the scope and press the ‘Print’ button. No set-up, nothing. The pictures generated are Bitmaps with 480 by 234 pixels.

Ever seen a heartbeat on a oscilloscope?

Ever seen a heartbeat on a oscilloscope?

The above picture shows the split screen feature of the LeCroy WaveAce 1002. Top: Full signal (Heartbeat), Bottom: Zoomed in on QRS complex

The above picture shows the split screen feature of the LeCroy WaveAce 1002. Top: Full signal (Heartbeat), Bottom: Zoomed in on QRS complex

If you have ever bought an inexpensive entry level DSO, you know that bandwidth is often an issue. The manufacturers of low-cost devices are really stretching the bandwidth data. The Owon PDS5022T for instance doesn’t even get close to its alleged 25 MHz. So how does the LeCroy measure up?

I wanted to measure the rise time of a pulse generator with my demo scope and determined the approximate bandwidth by dividing 0.35 by the rise time. It turns out that none of my signal generators are capable of putting out a signal with the needed rise time for 60 MHz (5.8 ns). But no worries, I will build a pulse generator soon and add the result here.

The actual rise time mentioned in the datasheet is actually a bit lower (5 ns). Therefore I am tempted to say that the 60 MHz rating is a rather conservative value with a bit of safety buffer.

UPDATE (10/24/2012): As promised, I measured the rise-time of the scope to determine the factual bandwidth. The rise time I measured is 3.6 ns, which is better than what the datasheet promises. Assuming a Gaussian response, the scope’s bandwidth would be close to 100 MHz (0.35 / 0.0000000036). For a flat response we’d get a bandwidth of around 120 MHz (0.44 / 0.0000000036). In either case, this is way beyond the 60 MHz the scope is rated for.

The actual rise time of the measured signal is around 1 ns. One can derive the scope's bandwidth mathematically from the measured rise time.

The actual rise time of the measured signal is around 1 ns. One can derive the scope’s bandwidth mathematically from the measured rise time.

With an antenna attached to Channel 1 input, I could capture a signal of my amateur radio transceiver on 146.520 MHz without a problem. Of course the time base doesn’t allow closer zoom on the signal and the amplitude is probably off. Nonetheless, it shows that the scope is definitely usable for a quick look at signals way beyond the oscilloscope’s bandwidth rating.

Even 145.520 MHz doesn't seem to be a problem for the 60 MHz version of the LeCroy WaveAce 1002 oscilloscope. The scope was loosely coupled to the signal source using an antenna.

Even 145.520 MHz doesn’t seem to be a problem for the 60 MHz version of the LeCroy WaveAce 1002 oscilloscope. The scope was loosely coupled to the signal source using an antenna.

Like all WaveAce oscilloscopes, the WaveAce 1002 has a bandwidth limit filter. If this feature is activated, the bandwith of the scope is limited to 20 MHz. By limiting the bandwidth, the noise that sometimes appears on the displayed waveform can be reduced. This results in a cleaner signal display.

The scope offers a variety of measurement parameters. The following parameters can be measured: Amplitude, Average, Base, Burst Width, Cyclical RMS, +Duty Cycle, -Duty Cycle, Fall Time, Frequency, Max, Mean, Min, Overshoot, Peak-Peak, Period, Phase, Preshoot, Rise Time, RMS, Top, +Width, -Width, First Rising-Rising, First Rising-Falling, First Falling-Rising, First Falling-Falling, Last Rising-Rising, Last Rising-Falling, Last Falling-Rising, Last Falling-Falling.

WaveAce 1002 with all measurements turned on. Looks a bit cluttered but with that amount of measurements to display, that's just how it's going to be.

WaveAce 1002 with all measurements turned on. Looks a bit cluttered but with that amount of measurements to display that’s just how it’s going to be.

An additional math track allows to display either the result of the mathematical functions add, subtract, multiply, and divide of channel 1 and 2. Additionally, a Fast Fourier Transform (FFT) of the signal on either channel 1 or 2 can be displayed.

With just 163 mm x 313 mm x 115.8 mm (6.42″ x 12.32″ x 4.6″) and a total weight of 2.78 kg (6.1 lbs), the scope is very compact. Thanks to its wide power supply range of 100 – 240 VAC and a frequency range of 45 Hz to 440 Hz, this scope is excellent for service all around the world. Thanks to the wide frequency range, the scope can even be used in special 400 Hz mains supply environments such as on board of an aircraft and in special military applications (MIL-STD-704).

The list price of the WaveAce 1002 is $990.00 [3]. For a real LeCroy, this is an excellent price / performance ratio.

Links and Sources:

[1] T&M Instruments: http://www.tandm.net/

[2] WaveAce 1000 and 2000 Oscilloscopes, LeCroy http://teledynelecroy.com/

[3] WaveAce 1002 Oscilloscope, LeCroy http://teledynelecroy.com/

Please cite this article as:
Westerhold, S. (2012), "Review: Teledyne LeCroy WaveAce 1002 – Digital Storage Oscilloscope". Baltic Lab High Frequency Projects Blog. ISSN (Online): 2751-8140., https://baltic-lab.com/2012/10/review-leccroy-waveace-1002-digital-storage-oscilloscope/, (accessed: November 23, 2024).

Funding:

If you liked this content, please consider contributing. Any help is greatly appreciated.

6 thoughts on “Review: Teledyne LeCroy WaveAce 1002 – Digital Storage Oscilloscope

  1. Did you measure the rise time with a probe? Please see my recent article in test and measurement world. The rule of thumb 0.95 is now closer to 1. We used the 640z in our example.

Leave a comment

Your email address will not be published. Required fields are marked *